Design of Limb Cooling Tourniquet to Prevent Amputation due to Ischemia

dc.contributor.advisorEmery, Ashley F.
dc.contributor.authorMhetre, Ketan Sunil
dc.date.accessioned2019-10-15T22:59:47Z
dc.date.available2019-10-15T22:59:47Z
dc.date.issued2019-10-15
dc.date.submitted2019
dc.descriptionThesis (Master's)--University of Washington, 2019
dc.description.abstractDuring a medevac response, tourniquets are often used as an effective device to mitigate blood loss in incidents involving serious injuries with bleeding. In emergent situations, often involving the transportation of the patient to a medical facility, the prevention of blood loss is prioritized by first responders. Thus, making the use of a tourniquet imperative. Although the device can prevent blood loss from the affected region, the process, however, leads to a deficiency in supplying oxygen which can often result in nerve and tissue damage. Several studies have shown that cooling the limb slows down the metabolism rate and might save a limb from possible amputation. This thesis discusses the process of developing a new limb cooling tourniquet, which consists of a cooling sleeve and an air splint coupled with a conventional tourniquet. To verify the efficiency of the newly developed cooling sleeve, a thermal phantom of a human leg was developed. Its performance was validated with the help of thermal simulation of human limb model. Animal limb cooling experiments were performed at Harborview Medical Center to study the effects of reduced blood flow and oxygen deficiency (ischemia) and the temperature profiles that were obtained as a result of this experiment were compared with the thermal simulation of an animal limb model and the simulations were validated.
dc.embargo.termsOpen Access
dc.format.mimetypeapplication/pdf
dc.identifier.otherMhetre_washington_0250O_20810.pdf
dc.identifier.urihttp://hdl.handle.net/1773/44855
dc.language.isoen_US
dc.rightsnone
dc.subjectAcute Limb Ischemia
dc.subjectLimb Amputation
dc.subjectMedical Devices
dc.subjectThermal Simulation of limb
dc.subjectTourniquet
dc.subjectMechanical engineering
dc.subject.otherMechanical engineering
dc.titleDesign of Limb Cooling Tourniquet to Prevent Amputation due to Ischemia
dc.typeThesis