Development of a Highly Sensitive Protein Diagnostic Device using Isothermal Strand Displacement Amplification

Loading...
Thumbnail Image

Authors

Guelig, Dylan

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

The lateral flow test (LFT) is the preferred tool for detection of protein biomarkers in low-resource settings. While low-cost and rapid, these immunoassay devices often suffer from limited sensitivity in the single chemical step format. Our group has addressed this deficiency by coupling a lateral flow test to isothermal strand displacement amplification (iSDA), a nucleic-acid amplification test (NAAT). This novel method is called LFT-iSDA and uses an oligonucleotide-labelled detection antibody in the immunoassay as the target template for iSDA. In its current embodiment, LFT-iSDA is ~10,000x more sensitive than gold nanoparticle-based LFTs, but is highly manual and relies on laboratory infrastructure. Here we present preliminary development of a diagnostic device integrating the highly sensitive LFT-iSDA into a cartridge supporting rapid time-to-results and minimal user-steps— a format amenable to decentralized test settings. We characterized performance limitations of this assay, developed a real-time fluorescence detection system, and used a two-dimensional paper network (2DPN) to semi-automate this multi-step chemical process. These results show important progress toward enabling highly sensitive protein detection in a user-friendly, low-cost device, at the point of care.

Description

Thesis (Master's)--University of Washington, 2018

Citation

DOI

Collections